INSTRUCTIONS FOR USE

RediCup device
The Reditest® RediCup® device is easy to use! Please follow the simple instructions below.

Instructions for testing of any combination of the following drugs: AMP, BAR, BZO, COC, THC, MTD, M-AMP, MDMA, MOP, OPI, OXY, PCP, TCA

A rapid, one step screening test for the simultaneous, qualitative detection of multiple drugs and drug metabolites in human urine.

NOTE: ALL POSITIVE RESULTS ARE PRESUMPTIVE AND SHOULD BE CONFIRMED BY AN ALTERNATIVE METHOD (e.g. GC/MS)

STEP 1

Collect specimen.
RediCup Step 1


STEP 2

Secure the cap. NOTE: Screw the cap tightly to ensure adequate seal.
RediCup Step 2


STEP 3

Quickly invert cup and place on a flat surface for approximately 30 seconds. NOTE: Do not swirl, tap or shake the cup as this can produce an invalid result.
RediCup Step 3


STEP 4

Return cup upright and wait 5 minutes.
RediCup Step 4


STEP 5

Read results at 5 minutes. Do not interpret results after 60 minutes as false results may occur. NOTE: Any line at all in the test zone is to be interpreted as negative.
RediCup Step 5



   

LABORATORY URINE DRUG TESTING

If test is positive, apply security seal over the lid and agency label with test(s) requested around the bottle. Send bottle and contents to laboratory for confirmation.
RediCup Step 6

Sending Presumptive Positive Specimens to the laboratory for GC/MS Confirmation

  1. Perform the on-site test according to the instructions provided above.
  2. Complete an Urine Drug Test Lab Requisition Form or Label to indicate which presumptive positive drugs you want to confirm through the lab. Make sure the donor signs the form. NOTE: If you wish to document preliminary positives and negatives for your own record, complete a Preliminary Result Form.
  3. Secure the positive specimen in the bottle. To prevent leakage, make sure the lid is threaded properly and tightly fastened.
  4. Place the security seal over the top of the RediCup.
  5. Fill out the Specimen ID/Chain of Custody Label and indicate the positive drug(s) on the line marked "GC/MS Confirm:___________".
  6. Place the label around the sides of the RediCup.
  7. Instruct the donor to initial the security seal.
  8. Seal the RediCup in a plastic baggie with the absorbent material.
  9. Include the “Lab Copy” of the Test Requisition Form in the shipping bag (five or more specimens) or the mailer box (fewer than five specimens) with the specimen.
  10. Send to RTL for GC/MS confirmation.

NOTE: If your account was not set up to include confirmation request forms, please contact your sales representative.

   

*PLEASE DO NOT SEND NEGATIVE SPECIMENS TO RTL FOR CONFIRMATION. NEGATIVE SPECIMENS WILL NOT BE CONFIRMED.